Our experienced engineers can design and manufacture custom fixtures for your specific sample profiles.
Controlled by Mecmesin’s powerful Emperor™ software
· Balanced to remove or apply axial compressive or tensile force
· Precise drive-to-loadcell alignment for extreme low torque testing
· Measurement from a few mN.m up to 6 N.m
The Helixa accepts interchangeable intelligent torque cells for measurement from just a few mN.m up to 6 N.m, with a resolution of 1:6,500 and an accuracy of ±0.5% of full scale.
Available also as a touch-screen model, the Helixa-xt, with all the same features but the convenience of a self-contained unit.
Specifications
· Intelligent torque cells rated at 100 mN.m, 200 mN.m, 300 mN.m, 1 N.m, 1.5 N.m, 3 N.m, and 6 N.m (14 ozf.in, 28 ozf.in, 42 ozf.in, 8.9 lbf.in, 13 lbf.in, 26.5 lbf.in, 53 lbf.in)
· Torque cell resolution: better than 0.01% of full scale
· Axial alignment: total runout (without fixtures), better than ±0.25 mm
· Speed from 0.1 to 30 rpm with resolution of ±0.1 rpm
· Sampling rates of 1 kHz, 500 Hz, 100 Hz, 50 Hz and 10 Hz
· Maximum displacement from tared position: 2500 revs
· Displacement accuracy: 0.1°
· Displayed resolution: 0.2°
· Throat depth (max sample radius): 111 mm, or 127 mm with bellows removed
· Headroom (without fixtures): 350 mm
· Dimensions: (with external weight hanger), Helixa-i: W 290 mm × H 758 mm × D 491 mm (Helixa-xt: W 587 mm)
· Weight: 28 kg (Helixa-i); 32 kg (Helixa-xt)